CounCell Penta 2.0

CounCell Penta 2.0

About CounCell

The name of the brand CounCell is derived from the counting of Whole Blood Cell. The brand supports the lab medicine fraternity by providing them with a state-of-the-art technology hematology analyzer in the continuously evolving In-vitro Diagnostic market for reliable and accurate blood cell counting.

Core Technology
  • CounCell Penta 2.0 utilizes the proven & reliable core technology of laser flow cytometry to count & analyze cells.
  • As cells are irradiated by the laser beam, the Tri-angle scattered light of the cell are analyzed.
  • The 3 signals used to differentiate & count cells with the help of unique digital technology & algorithm.
banner
Results
  • With 36 Parameters, 4 Scattergrams & 2 histograms, the instrument gives output based on Tri-angle laser scatter and Electrical impedance method.
Dependable and reliable software
  • Easy and comprehensive maintenance and system status monitoring.
  • Smart prompt and one touch error resolving capabilities.
Reliable and accurate results
  • Independent Basophil measurement..
  • Alarm Flag for better diagnostics..
  • Tri-level QC: I-J and X-B mode..
  • Wider linearity range.
Customised user friendly system
  • Multiple analysis mode: VWB, Capillary and PD (with CBC and CBC + DIFF).
  • 10.4 inch colour touch screen.
  • Real time reagent management.
  • Built in thermal printer.

Technical Specifications

Sr. No. Parameter Description
1. Principle Flow Cytometry + Triangle laser scatter method for WBC and its Differential counting. Electrical impedance method for RBC/PLT and Cyanide Free Colorimetric method for HGB.
2. Reportable Parameters 36 Parameters consisting of 8 RUO parameters:
WBC, LYM%, MON%, NEU%, BAS%, EOS%, LYM#, MON#, NEU#, EOS#, BAS#, NLR, PLR, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW-CV, PDW-SD, PCT, P-LCR, P-LCC, NRBC%, NRBC#, PLT CLUMPS,%%, PLT#, LIC%, LIC#, ALY%, ALY#.
4-Scattergrams 2-Histograms for RBC and PLT
Alarm for abnormal erythrocytes, leukocytes and platelet.
3. Throughput 60 Samples per hour.
4. Reagent 3 Routine reagents: Diluent & LH Lyse & Diff Lyse
5. Sample Volume 20µL
6. Data Storage Capacity Internal Memory can store 600,000 test values with parameters, Scattergram, Histograms and Patient information.
7. Quality Control 2 QC Method: L-J and X-B QC with appropriate QC charts.
8. Calibration 3 Types of Calibration, Manual Calibration, Auto Calibration with calibrators and Auto Calibration using fresh sample.
9. Power Requirement 100v-240 VAC ; 50/60Hz.
10. Operating Environment Temperature: 10°C 30°C,
Relative Humidity: 30% - 85%,
Atmospheric pressure: 70kPa 106kPa.
11. Dimension 430 mm(L) X 350mm(W) X 450mm(H)
12. Weight 28kg

Linearity & Repeatability

Sr. No. Parameters Linearity Repeatability
1. WBC (0.0 - 300.0) X 109/L < 2.0%
2. RBC (0.0 - 8.0) X 1012/L < 1.5%
3. HGB (0.0 - 25) g/dL < 1.5%
4. PLT (0 - 5000.0) X 109/L < 4.0%