| 1 |
INTERMEDIATE PRECISION |
%CV at Normal Control Level<=7.1 |
%CV=2.9 (at Normal Control) |
| %CV at Pathological Control <=7.1 (WHO: Laboratory Diagnosis and Monitoring of Diabetes Mellitus, 2002) |
%CV=2.6 (at Pathological Control) |
| REPEATABILITY |
%CV at 51-110mg/dL glucose<=7.1 |
%CV=2.6 (at 51-110mg/dL glucose) |
| %CV at 111-150mg/dL glucose<=7.1 |
%CV=2.7 (at 111-150mg/dL glucose) |
| 2 |
SYSTEM ACCURACY AND BIAS EVALUATION |
%CV at 151-250mg/dL glucose<=7.1 (WHO: laboratory Diagnosis and Monitoring of Diabetes Mellitus, 2002) |
%CV=2.8 (at 151-250mg/dL glucose) |
| System Accuracy requirement: 95% of the individual Glucometer results shall fall within +/- 15mg/dL of the results of the 'Reference Method' at glucose concentrations <100mg/dL and within +/- 15% at glucose concentrations >=100mg/dL (**ISO15197 :2013**) |
100% of individual glucometer results are within +/- 15mg/dL of the results of the 'Reference Method' at glucose concentration <100mg/dL
100% of individual glucometer results are within +/- 15% of the results of the 'Reference Method' at glucose concentration >=100mg/dL |
| Maximum allowable deviation should be <= 15% (WHO: Laboratory Diagnosis and Monitoring of Diabetes Mellitus, 2002) |
% Inaccuracy obtained-3.4 (lower limit) and-0.8(Upper limit) at 95% confidence interval. |