CounCell 23 Excel

CounCell 23 Excel

CounCell 23 Excel - 3-Diff Auto Hematology Analyser

CounCell 23 Excel 3-Diff Auto Hematology Analyzer is specially designed in accordance to international standards. The system has state of art technology, customised user friendly software and robust hardware. The system is based on principle of Electrical impedance for counting of WBC, RBC & PLT and Cyanide free colorimetric method for HGB estimation.

Dependable and reliable technology:
  • Electrical Impedance method for WBC/RBC/PLT counting.
  • Cyanide free colorimetric method for HGB estimation.
  • The system uses a special counting mode to give precise values in clinically important WBC and PLT values.
banner
Powerful software:
  • Easy and comprehensive maintenance and system status monitoring.
  • Smart prompt and one-touch error resolving capabilities.
  • Real-time reagent management.
Reliable and accurate results:
  • Tri-level QC, L-J and X-B mode.
  • The Multiple calibration mode: Automatic calibration with Calibrator/Fresh blood sample and manual calibration.
  • Wider Linearity range.
  • Multiple analysis mode: VVB, Capillary and PD.
Customised user friendly system
  • 10.4 inch colour touch screen.
  • Multiple user login as per user request.
  • Built-in thermal printer.
  • USB and 1 network interface.

Technical Specifications

Sr. No. Parameter Description
1. Principle Electrical impedance method for WBC/RBC/PLT. Cyanide Free Colorimetric method for HGB.
2. Reportable Parameters 24 Parameters Consisting:
WBC, LYM#, LYM%, MID#, MID%, GRAN#, GRAN%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, MPV, PCT, PDW-SD, PDW-CV, P-LCC, P-LCR, NLR & PLR.
3 Histograms for WBC, RBC & PLT.
3. Throughput 60 Samples per hour.
4. Reagent 2 Routine Reagents: Diluent & Lyse.
5. Sample Volume 9µL
6. Data Storage Capacity Internal Memory can store 600,000 test values with parameters, Histograms and Patient information.
7. Quality Control 2 QC Method; L-J and X-B QC with appropriate QC charts.
8. Calibration 3 Types of Calibration, Manual Calibration, Auto Calibration with calibrators and Auto Calibration using fresh sample.
9. Power Requirement 100v-240 VAC; 50/60Hz.
10. Operating Environment Temperature: 10°C - 30°C,
Relative Humidity: 30% - 85%,
Atmospheric pressure: 70kPa - 106kPa.
11. Dimension 430 mm(L) X 280mm(W) X 400mm(H)
12. Weight 17.5 Kg

Linearity & Repeatability

Parameters Linearity Repeatability
WBC (0.0 - 300.0) X 109/L ≤ 2.0%
RBC (0.0 - 8.0) X 1012/L ≤ 1.5%
HGB (0.0 - 25) g/dL ≤ 1.5%
PLT (0.0 - 5000.0) X 109/L ≤ 4.0%